Melbourne, FL Certification

View PDF

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017, ANSI/NCSL Z540-1-1994 (R2002) AND ANSI/NCSL Z540.3-2006 (R2013)

Technical Maintenance, Inc.
3060 Venture Lane
Melbourne, FL 32934
Scott Chamberlain 321-242-0890

CALIBRATION

Valid to: September 20, 2019 Certificate Number: AC-2080.08

Chemical Quantities

Parameter/Equipment Range Expanded Uncertainty of

Measurement (+/-)

Reference Standard, Method, and/or Equipment
pH1 (4, 7, 10) pH 0.02 pH pH buffer solutions
Conductivity, Liquid1 (2 to 10 000) µS 1 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading Conductivity solutions

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of

Measurement (+/-)

Reference Standard, Method, and/or Equipment
DC Current – Generate1 Up to 220 µA

220 µA to 2.2 mA (2.2 to 22) mA (22 to 220) mA

220 mA to 2.2 A

40 µA/A + 6 nA

35 µA/A + 7 nA

36 µA/A + 40 nA

48 µA/A + 0.7 µA

81 µA/A + 12 µA

Fluke 5730A/03

Multifunction Calibrator

DC Current – Generate1 (2.2 to 11) A 0.36 mA/A + 0.48 mA Fluke 5730A/03

Multifunction Calibrator

5725A Amplifier

DC Current – Generate1 (11 to 20.5) A 0.76 mA/A + 0.75 mA Fluke 5522A Multifunction Calibrator
DC Current – Generate1

Clamp Only

(20.5 to 1 000) A 0.9 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading Coil/ Fluke 5522A Multifunction Calibrator



Version 004 Issued: November 10, 2017 www.anab.org

Page 1 of 12

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of

Measurement (+/-)

Reference Standard, Method, and/or Equipment
DC Current – Measure1 Up to 220 µA

220 µA to 2.2 mA (2.2 to 22) mA (22 to 220) mA (0.22 to 1) A

6 µA/A

5 µA/A

6 µA/A

10 µA/A

37 µA/A

Agilent 3458A/002

Precision Multimeter

DC Voltage1 – Generate (0 to 220) mV

220 mV to 2.2 V (2.2 to 11) V (11 to 22) V (22 to 220) V

(220 to 1 100) V

11 µV/V + 0.4 µV

5.2 µV/V + 0.7 µV

3.5 µV/V + 2.5 µV

3.5 µV/V + 4 µV

5 µV/V + 40 µV

6.5 µV/V + 400 µV

Fluke 5730A/03

Multifunction Calibrator

DC Voltage1 – Measure Up to 100 mV (0.1 to 1) V (1 to 10) V

(10 to 100) V (100 to 1 000) V

10 µV/V + 0.3 µV

9.0 µV/V + 0.30 µV

9.0 µV/V + 50 µV

11 µV/V + 30 µV

11 µV/V + 0.1 mV

Agilent 3458A/002

Precision Multimeter

DC Voltage1 – Measure (1 to 120) kV 0.23 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading Ross VD120 High

Voltage Divider

HP 34401A Precision

Multimeter

AC Voltage – Generate1 (0.22 to 2.2) mV

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz

500 kHz to 1 MHz

(2.2 to 22) mV

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz

500 kHz to 1 MHz

0.27 mV/V + 4 µV

0.14 mV/V + 4 µV

0.13 mV/V + 4 µV

0.22 mV/V + 4 µV

0.52 mV/V + 5 µV

1. 1 mV/V + 10 µV

1.4 mV/V + 20 µV

2.7 mV/V + 20 µV

0.24 mV/V + 4 µV

91 µV/V + 4 µV

81 µV/V + 4 µV

0.2 mV/V + 4 µV

0.5 mV/V + 5 µV

1.1 mV/V + 10 µV

1.4 mV/V + 20 µV

2.7 mV/V + 20 µV

Fluke 5730A/03

Multifunction Calibrator

Version 004 Issued: November 10, 2017 www.anab.org

Page 2 of 12

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of

Measurement (+/-)

Reference Standard, Method, and/or Equipment
AC Voltage – Generate1 (22 to 220) mV

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz

500 kHz to 1 MHz

220 mV to 2.2 V (10 to 20) Hz (20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz

(50 to 100) kHz (100 to 300) kHz (300 to 500) kHz

500 kHz to 1 MHz

(2.2 to 22) V

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz

500 kHz to 1 MHz

(22 to 220) V

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz

(50 to 100) kHz

(220 to 750) V

40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz

0.24 mV/V + 12 µV

90 µV/V + 7 µV

58 µV/V + 7 µV

0.12 mV/V + 7 µV

0.31 mV/V + 17 µV

0.66 mV/V + 20 µV

1.4 mV /V + 25 µV

2.7 mV /V + 45 µV

0.24 mV/V + 40 µV

90 µV/V + 15 µV

43 µV/V + 8 µV

67 µV/V + 10 µV

85 µV/V + 30 µV

0.34 mV/V + 80 µV

1 mV /V + 0.2 mV

1.7 mV/V + 0.3 mV

0.24 mV/V + 0.4 mV

91 µV/V + 0.15 mV

43 µV/V + 50 µV

67 µV/V + 0.1 mV

83 µV/V + 0.2 mV

0.25 mV/V + 0.6 mV

1 mV/V + 2 mV

1 5 mV/V + 3.2 mV

0.24 mV /V + 4 mV

90 µV/V + 1.5 mV

52 µV/V + 0.6 mV

80 µV/V + 1 mV

0.15 mV /V + 2.5 mV

91 µV/V + 4 mV

0.17 mV /V + 6 mV

0.6 mV /V + 11 mV

2 3 mV /V + 45 mV

Fluke 5730A/03

Multifunction Calibrator

Version 004 Issued: November 10, 2017 www.anab.org

Page 3 of 12

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of

Measurement (+/-)

Reference Standard, Method, and/or Equipment
AC Voltage – Generate1 (750 to 1 000) V

40 Hz to 1 kHz

(1 to 20) kHz

(20 to 30) kHz

91 µV/V + 4 mV

0.17 mV /V + 6 mV

0.6 mV /V + 11 mV

Fluke 5730A/03

Multifunction Calibrator

5725A Amplifier

AC Voltage – Measure1 Up to 10 mV

(1 to 40) Hz

40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz

(100 to 300) kHz

10 mV to 10 V

(1 to 40) Hz

40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz

(100 to 300) kHz

300 kHz to 1 MHz

(1 to 2) MHz

(10 to 100) V

(1 to 40) Hz

40 Hz to 1 kHz

(1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz

300 kHz to 1 MHz

0.03 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.03 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.02 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.011 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.03 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.011 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.1 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.011 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.011 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

4 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.02 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.007 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.004 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.007 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.014 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.03 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.8 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.3 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.01 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

1 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.01 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

1.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.01 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.02 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.004 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.02 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.02 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.04 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.12 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.4 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

1.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.01 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

Agilent 3458A/002

Precision Multimeter

AC Voltage – Measure1 (100 to 1 000) V

(1 to 40) Hz

40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz

0.05 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.004 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.04 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.06 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.12 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

0.3 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of range

Agilent 3458A/002

Precision Multimeter

AC Voltage – Measure1 (1 to 85) kV

(50, 60) Hz

0.7 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading Ross VD120 High

Voltage Divider

HP 34401A Precision

Multimeter

Version 004 Issued: November 10, 2017 www.anab.org

Page 4 of 12

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of

Measurement (+/-)

Reference Standard, Method, and/or Equipment
AC Current – Generate1 (9 to 220) µA

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 1 kHz

(1 to 5) kHz

(5 to 10) kHz

0.25 mA/A + 16 nA

0.16 mA /A + 10 nA

0.1 mA /A + 8 nA

0.28 mA /A + 12 nA

1.1 mA /A + 65 nA

Fluke 5730A/03

Multifunction Calibrator

AC Current – Generate1 (0.22 to 2.2) mA

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 1 kHz

(1 to 5) kHz

(5 to 10) kHz

(2.2 to 22) mA

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 1 kHz

1 to 5 kHz

(5 to 10) kHz

(22 to 220) mA

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 1 kHz

(1 to 5) kHz

(5 to 10) kHz

0.25 mA /A + 40 nA

0.16 mA /A + 35 nA

0.1 mA /A + 35 nA

0.2 mA /A + 0.110 µA

1.1 mA /A + 0.650 µA

0.25 mA /A + 0.4 µA

0.16 mA /A + 0.35 µA

0.1 mA /A + 0.35 µA

0.2 mA /A + 0.55 µA

1.1 mA /A + 5 µA

0.25 mA /A + 4 µA

0.16 mA /A + 3.5 µA

0.1 mA /A + 2.5 µA

0.2 mA /A + 3.5 µA

1.1 mA /A + 10 µA

Fluke 5730A/03

Multifunction Calibrator

AC Current – Generate1 (0.22 A to 2.2) A

20 Hz to 1 kHz

(1 to 5) kHz

(5 to 10) kHz

0.25 mA/A + 35 µA

0.45 mA /A + 80 µA

7 mA /A + 0.16 mA

Fluke 5730A/03

Multifunction Calibrator

5725A Amplifier

AC Current – Generate1 (2.2 to 11) A

40 Hz to 1 kHz

(1 to 5) kHz

(5 to 10) kHz

(11 to 20.5) A

(45 to 100) Hz

100 Hz to 1 kHz

(1 to 5) kHz

0.46 mA /A + 0.17 mA

0.95 mA /A + 0.38 mA

3.6 mA /A + 0.75 mA

0.1 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 2 mA

0.12 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 5 mA

2.3 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 5 mA

Fluke 5522A Multifunction Calibrator
AC Current – Generate1

Clamp Only

(20.5 to 500) A

(45 to 440) Hz

1.8 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading Coil / Fluke 5522A Multifunction Calibrator

Version 004 Issued: November 10, 2017 www.anab.org

Page 5 of 12

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of

Measurement (+/-)

Reference Standard, Method, and/or Equipment
AC Current – Measure1 Up to 100 µA

(10 to 20) Hz

(20 to 45) Hz

45 Hz to 1 kHz

0.41 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.03 pA

0.16 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.03 pA

0.07 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.03 pA

Agilent 3458A/002

Precision Multimeter

AC Current – Measure1 (0.1 to 100) mA

(10 to 20) Hz (20 to 45) Hz (45 to 100) Hz

100 Hz to 5 kHz

(5 to 20) kHz

(20 to 50) kHz

(50 to 100) kHz

(0.1 to 1) A

(10 to 20) Hz (20 to 45) Hz (45 to 100) Hz

100 Hz to 5 kHz

(5 to 20) kHz

(20 to 50) kHz

0.41 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 20 μA

0.16 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 20 μA

0.07 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 20 μA

0.04 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 20 μA

0.08 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 20 µA

0.41 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 40 µA

0.56 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 150 µA

0.41 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mA

0.17 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mA

0.09 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mA

0.11 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mA

0.31 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mA

1 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.4 mA

Agilent 3458A/002

Precision Multimeter

Resistance – Generate, Fixed Points1 1, 1.9 Ω

10, 19 Ω

100, 190 Ω

1, 1.9 kΩ

10, 19 kΩ

100, 190 kΩ

1 MΩ

1.9 MΩ

10 MΩ

19 MΩ

100 MΩ

98 μΩ/Ω

23 μΩ/Ω

10 μΩ/Ω

6.5 μΩ/Ω

6.5 μΩ/Ω

8.5 μΩ/Ω

13 μΩ/Ω

18 μΩ/Ω

40 μΩ/Ω

48 μΩ/Ω

0.1 mΩ/Ω

Fluke 5730A/03

Multifunction Calibrator

Version 004 Issued: November 10, 2017 www.anab.org

Page 6 of 12

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of

Measurement (+/-)

Reference Standard, Method, and/or Equipment
Resistance – Generate1 Up to 11 Ω (11 to 33) Ω (33 to 110) Ω (110 to 330) Ω

330 Ω to 1.1 kΩ

(1.1 to 3.3) kΩ (3.3 to 11) kΩ (11 to 33) kΩ (33 to 110) kΩ (110 to 330) kΩ

(0.33 to 1.1) MΩ

(1.1 to 3.3) MΩ (3.3 to 11) MΩ (11 to 33) MΩ (33 to 110) MΩ (110 to 330) MΩ

330 MΩ to 1.1 GΩ

40 μΩ/Ω + 0.01 Ω

30 μΩ/Ω + 0.015 Ω

28 μΩ/Ω + 0.015 Ω

28 μΩ/Ω + 0.02 Ω

29 μΩ/Ω + 0.02 Ω

29 μΩ/Ω + 0.2 Ω

29 μΩ/Ω + 0.1 Ω

29 μΩ/Ω + 1 Ω

29 μΩ/Ω + 1 Ω

33 μΩ/Ω + 10 Ω

33 μΩ/Ω + 10 Ω

60 μΩ/Ω + 150 Ω

0.013 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.25 kΩ

0.025 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 2.5 kΩ

0.05 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 3.0 kΩ

0.3 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 100 kΩ

1.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 500 kΩ

Fluke 5522A Multifunction Calibrator
Resistance – Measure1

Fixed Points

Up to 10 Ω

(10 to 100) Ω

100 Ω to 100 kΩ

100 kΩ to 1 MΩ

(1 to 10) MΩ

(10 to 100) MΩ

100 MΩ to 1 GΩ

16 μΩ/Ω + 50 µΩ

14 μΩ/Ω + 0.5 mΩ

12 μΩ/Ω + 5 mΩ

16 μΩ/Ω + 2 Ω

50 μΩ/Ω + 100 Ω

0.05{3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 1 kΩ

9.2 μΩ/Ω + 1 Ω

Agilent 3458A/002

Precision Multimeter

Capacitance – Generate1 (0.19 to 0.1099) nF (1.1 to 3.299) nF (3.3 to 329.999) nF

(0.33 to 1.09999) µF (1.1 to 3.29999) µF

(3.3 to 10.9999) µF (11 to 32.999) µF

(33 to 109.999) µF (110 to 329.999) µF (0.33 to 1.09999) mF

(1.1 to 3.2999) mF (3.3 to 10.9999) mF

(11 to 32.9999) mF (33 to 110) mF

1.2 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.01 nF

1.2 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.01 nF

0.39 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.3 nF

0.38 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 3 nF

0.37 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 3 nF

0.38 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 10 nF

0.53 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 30 nF

0.58 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.1 µF

0.58 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.3 µF

0.57 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 1 µF

0.58 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 3 µF

0.59 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 10 µF

0.65 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 30 µF

1.3 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.1 mF

Fluke 5522A Multifunction Calibrator

Version 004 Issued: November 10, 2017 www.anab.org

Page 7 of 12

Electrical – DC/Low Frequency

Parameter/Equipment Range Expanded Uncertainty of

Measurement (+/-)

Reference Standard, Method, and/or Equipment
Oscilloscope Calibration – Vertical Deflection1

50 Ω, 1 kHz square wave

1 MΩ, 1 kHz square wave

Flatness – Leveled Sine Wave

Bandwidth

Rise Time

Time Interval2

1 mV to 6.6 V

1 mV to 130 V

5 mV to 5.5 V

50 kHz to 100 MHz (100 to 300) MHz (300 to 600) MHz (600 to 1 100) MHz

<300 ps

1 ns to 20 ms

50 ms to 5 s

0.59 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 40 µV

0.44 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 40 µV

2.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.1 mV

2.2 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.1 mV

4.6 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.1 mV

6.4 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.1 mV

+ 13 ps / -110 ps

3 μs/s

(26 + 1 000t) μs/s

Fluke 5522A/SC1100

Multifunction Calibrator

Power Meter Range

Calibration1

3 μW to 100 mW 0.25 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading HP 11683A Power Meter Calibrator
Electrical Calibration of Thermocouple Indicating Devices1 Type B

(600 to 1 820) °C Type C

(0 to 2 316) °C Type E

(-250 to 1 000) °C

Type J

(-210 to 1 200) °C Type K

(-200 to 1 372) °C

Type L

(-200 to 900) °C Type N

(-200 to 1 300) °C Type R

(0 to 1 767) °C

Type S

(0 to 1 767) °C

0.41 °C

0.38 °C

0.29 °C

0.19 °C

0.21 °C

0.12 °C

0.26 °C

0.45 °C

0.44 °C

Fluke 7526A Process Calibrator

 

Thermodynamic

Parameter/Equipment Range Expanded Uncertainty of

Measurement (+/-)

Reference Standard, Method, and/or Equipment
Relative Humidity – Measure (0 to 90) {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443}RH 1.7 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443}RH Vaisala MI70 / HMP75

Thermohygrometer

Temperature – Measuring Equipment1 (-25 to 350) °C 0.12 °C Hart 1502A with 5616 PRT

and dry block

Temperature – Measure1 (-25 to 600) °C 0.025 °C Hart 1502A with 5616 PRT
IR Thermometry1 (20 to 100) ºC (100 to 300) ºC (300 to 500) ºC 0.51 ºC

0.61 ºC

0.8 ºC

Fluke 9132 Blackbody

ε = 0.95, λ= (8 to 14) μm

Time and Frequency

Parameter/Equipment Range Expanded Uncertainty of

Measurement (+/-)

Reference Standard, Method, and/or Equipment
Frequency – Measure 10 MHz 1 x 10-11 Hz/Hz HP 58503A/B GPS
Frequency – Measure1 10 Hz to 26.5 GHz 1 x 10-8 Hz/Hz HP 53131A Counter, Agilent E4440A Spectrum Analyzer

Calibration and Measurement Capability (CMC) is expressed in terms of the measurement parameter, measurement range, expanded uncertainty of measurement and reference standard, method, and/or equipment. The expanded uncertainty of measurement is expressed as the standard uncertainty of the measurement multiplied by a coverage factor of 2 (k=2), corresponding to a confidence level of approximately 95{3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443}.

Notes:

1. On-site calibration service is available for this parameter, since on-site conditions are typically more variable than those in the laboratory, larger measurement uncertainties are expected on-site than what is reported on the accredited scope

2. L = length in inches, t = time in seconds, D = diameter in inches, R = resolution of the device under test

3. This scope is formatted as part of a single document including Certificate of Accreditation No. AC-2080.08.
Version 004 Issued: November 10, 2017 www.anab.org