SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017, ANSI/NCSL Z540-1-1994, AND ANSI/NCSL Z540.3-2006
Technical Maintenance, Inc.
3000 Northwoods Parkway, Suite 270
Peachtree Corners, GA 30071
Scott Chamberlain Phone: 770-409-8348
CALIBRATION
Valid to: September 20, 2018 Certificate Number: AC-2080.01
Chemical
Parameter/ Equipment |
Range |
Calibration and Measurement Capability [Expressed as Uncertainty (±)] |
Reference Standard or Equipment |
pH – Meters3 |
(4, 7, 10) pH |
0.02 pH |
Standard pH buffers |
Dimensional
Parameter/ Equipment |
Range |
Calibration and Measurement Capability [Expressed as Uncertainty (±)] |
Reference Standard or Equipment |
Calipers2,3 |
Up to 46 in |
(47 + 8L) µin |
Grade 2 gage blocks |
Micrometers2,3 |
Up to 46 in |
(28 + 5L) µin |
Grade 2 gage blocks |
Bore Micrometers |
(0.125 to 2) in |
98 µin |
Master ring gages |
Dial Indicators2,3 |
Up to 0.2 in Up to 6 in Up to 1 in |
15 µin (81 + 1.2L) µin 150 µin |
Grade 2 gage blocks Starrett 716 |
Height Gages2,3 |
Up to 46 in |
(300 + 2L) µin |
Grade 2 gage blocks |
Scales – Rulers3 |
Up to 46 in |
0.009 in |
Grade 2 gage blocks |
Feeler Gages3 |
Up to 1 in |
93 µin |
Mitutoyo 293-369 |
Surface Plates3 – Overall Flatness Local Area Flatness |
(18 x 18) in to (6 × 6) ft Up to (6×6) ft |
170 µin 120 µin |
Rahn Planekator Repeat-o-meter |
Version 001 Issued: 10/14/2016 Page 1 of 11
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org
Dimensional
Parameter/ Equipment |
Range |
Calibration and Measurement Capability [Expressed as Uncertainty (±)] |
Reference Standard or Equipment |
Gage Blocks2 |
Up to 10 in |
(3.3 + 2.5L) μin |
Universal measuring machine, master gage block set |
Protractors3 |
(0 to 360) º |
0.014 º |
Angle blocks |
Radius Gages |
(0.125 to 1) inch |
340 µin |
Optical comparator |
Cylindrical Gages2,3 – Plain Pin, Plugs Rings |
(0 to 12) in (0.04 to 14) in |
(5.4 + 2.2D) µin (16 + 1.5D) µin |
Master gage blocks, P&W universal measuring machine |
Thread Plugs – Major Pitch Diameter (6 to 80) TPI |
Up to 12 in Up to 12 in |
36 µin 91 µin |
Gage blocks, P & W universal measuring machine, VanKeren thread wire set |
Thread Rings2 |
Up to 12 in |
(350 + 47D) µin |
Thread setting plug gages |
Dimensional Testing
Parameter/ Equipment |
Range |
Calibration and Measurement Capability [Expressed as Uncertainty (±)] |
Reference Standard or Equipment |
1D Length |
Up to 12 in |
350 µin |
Optical comparator |
Angle |
Up to 360º |
0.06º |
Optical comparator |
Electrical – DC/Low Frequency
Parameter/ Equipment |
Range |
Calibration and Measurement Capability [Expressed as Uncertainty (±)] |
Reference Standard or Equipment |
DC Voltage – Generate3 |
Up to 220 mV (0.22 to 2.2) V (2.2 to 11) V (11 to 22) V (22 to 220) V (220 to 1 100) V |
11 µV/V + 0.4 µV 8.2 µV/V + 0.7 µV 7.1 µV/V + 2.5 µV 7.1 µV/V + 4 µV 8.2 µV/V + 40 µV 10 µV/V + 0.4 mV |
Fluke 5720A |
DC Voltage – Generate3 |
(1 to 60) kV |
0.13 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} |
Ross VD60 |
DC Voltage – Measure3 |
Up to 100 mV (0.1to 1) V (1 to 10) V (10 to 100) V (100 to 1 000) V |
10 μV/V + 0.3 μV 9 μV/V + 0.3 μV 9 μV/V + 0.5 μV 11 μV/V + 30 μV 11 μV/V + 0.1 mV |
HP 3458A |
DC Voltage – Measure3 |
(1 to 6) kV |
2.3 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 10 V |
Associated Research 5560DT |
Version 001 Issued: 10/14/2016 Page 2 of 11
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org
Electrical – DC/Low Frequency
Parameter/ Equipment |
Range |
Calibration and Measurement Capability [Expressed as Uncertainty (±)] |
Reference Standard or Equipment |
DC Current – Generate3 |
Up to 220 µA (0.22 to 2.2) mA (2.2 to 22) mA (22 to 220) mA (0.22 to 2.2) A |
55 µA/A + 6 nA 50 µA/A + 7 nA 50 µA/A + 40 nA 60 µA/A + 0.7 µA 98 µA/A + 12 µA |
Fluke 5720A |
DC Current – Generate3 |
(2.2 to 20.5) A |
0.11 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 750 µA |
Fluke 5522A |
DC Current – Generate3Clamp Only |
Up to 1 000 A |
0.77 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.5 A |
Fluke 5522A / coil |
DC Current – Measure3 |
Up to 100 nA (0.1 to 1) µA (1 to 10) µA (10 to 100) µA (0.1to 10) mA (10 to 100) mA (0.1 to 1) A |
36 μA/A + 0.04 nA 24 μA/A + 0.04 nA 24 μA/A + 0.1 nA 24 μA/A + 0.8 nA 25 μA/A + 0.05 μA 43 μA/A + 0.5 μA 0.14 mA/A + 10 μA |
HP 3458A |
DC Current – Measure3 |
(1 to 500) A |
0.32 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} |
Current shunts |
Resistance3 – Fixed Points |
1 Ω, 1.9 Ω 10 Ω, 19 Ω 100 Ω, 190 Ω 1 kΩ, 1.9 kΩ 10 kΩ, 19 kΩ 100 kΩ, 190 kΩ 1 MΩ 1.9 MΩ 10 MΩ 19 MΩ 100 MΩ |
0.12 mΩ/Ω 31 μΩ/Ω 23 μΩ/Ω 12 μΩ/Ω 13 μΩ/Ω 14 μΩ/Ω 24 μΩ/Ω 26 μΩ/Ω 50 μΩ/Ω 59 μΩ/Ω 0.14 mΩ/Ω |
Fluke 5720A |
Resistance3 – Generate |
Up to 11 Ω (11 to 33) Ω (33 to 110) Ω (110 to 330) Ω (0.33 to 1.1) kΩ (1.1 to 3.3) kΩ (3.3 to 11) kΩ (11 to 33) kΩ (33 to 110) kΩ (110 to 330) kΩ (0.33 to 1.1) MΩ (1.1 to 3.3) MΩ (3.3 to 11) MΩ (11 to 33) MΩ (33 to 110) MΩ (110 to 330) MΩ (0.33 to 1.1) GΩ |
47 μΩ/Ω + 10 mΩ 42 μΩ/Ω + 15 mΩ 34 μΩ/Ω + 15 mΩ 32 μΩ/Ω + 0.02 Ω 33 μΩ/Ω + 0.02 Ω 33 μΩ/Ω + 0.1 Ω 33 μΩ/Ω + 0.1 Ω 33 μΩ/Ω + 1 Ω 33 μΩ/Ω + 1 Ω 37 μΩ/Ω + 10 Ω 37 μΩ/Ω + 10 Ω 70 μΩ/Ω + 0.15 kΩ 0.015 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.25 kΩ 0.029 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 2.5 kΩ 0.06 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 3 kΩ 0.35 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.1 MΩ 1.7 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.5 MΩ |
Fluke 5522A |
Resistance – Measure3 |
Up to 10 Ω (10 to 100) Ω (0.1 to 100) kΩ (0.1 to 1) M (1 to 10) MΩ (10 to 100) MΩ (0.1 to 1) GΩ |
18 μΩ/Ω + 5 μΩ 15 μΩ/Ω + 5 μΩ 13 μΩ/Ω + 5 mΩ 18 μΩ/Ω + 2 Ω 58 μΩ/Ω + 0.1 kΩ 0.059 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 1 kΩ 0.29 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 10 kΩ |
HP 3458A |
Version 001 Issued: 10/14/2016 Page 3 of 11
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org
Electrical – DC/Low Frequency
Parameter/ Equipment |
Range |
Calibration and Measurement Capability [Expressed as Uncertainty (±)] |
Reference Standard or Equipment |
Capacitance – Generate3 |
(0.19 to 0.1099) nF (1.1 to 3.299) nF (3.3 to 329.999) nF (0.33 to 3.29999) μF (3.3 to 10.999) μF (10.999 to 32.9999) μF (33 to 109.999) μF (110 to 329.999) μF (0.33 to 1.09999) mF (1.1 to 3.2999) mF (3.3 to 10.9999) mF (11 to 32.9999) mF (33 to 110) mF |
0.6 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.01 nF 0.32 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.01 nF 0.32 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.3 nF 0.32 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 3 nF 0.32 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 10 nF 0.48 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 30 nF 0.54 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.1 µF 0.55 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.3 µF 0.54 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 1 μF 0.54 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 3 μF 0.54 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 10 μF 0.87 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 30 μF 1.3 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 100 μF |
Fluke 5522A |
Capacitance – Measure3 |
0.1 pF to 10 mF |
0.78 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of rdg |
Fluke PM 6304C |
AC Voltage – Generate3 |
Up to 22 mV (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz 500 kHz to 1 MHz (22 to 220) mV (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz (0.22 500 kHz to 1 MHz to 2.2) V (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz 500 kHz to 1 MHz (2.2 to 22) V (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz 500 kHz to 1 MHz |
0.36 mV/V + 5 µV 0.15 mV/V + 5 µV 0.13 mV/V + 5 µV 0.3 mV/V + 5 µV 0.71 mV/V + 6 µV 1.6 mV/V + 12 µV 2 mV/V + 25 µV 4 mV/V + 25 µV 0.35 mV/V + 15 µV 0.14 mV/V + 8 µV 0.12 mV/V + 8 µV 0.29 mV/V + 8 µV 0.7 mV/V + 20 µV 1.3 mV/V + 25 µV 2 mV/V + 30 µV 3.9 mV/V + 60 µV 0.51 mV/V + 50 µV 0.4 mV/V + 20 µV 0.38 mV/V + 10 µV 1.4 mV/V + 12 µV 0.41 mV/V + 40 µV 0.69 mV/V + 0.1 mV 1.5 mV/V + 0.25 mV 2.4 mV/V + 0.4 mV 0.48 mV/V + 0.5 mV 0.39 mV/V + 0.2 mV 0.38 mV/V + 70 µV 0.39 mV/V + 0.12 mV 0.4 mV/V + 0.25 mV 0.5 mV/V + 0.8 mV 1.3 mV/V + 2.5 mV 1.9 mV/V + 4 mV |
Fluke 5720A |
Version 001 Issued: 10/14/2016 Page 4 of 11
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org
Electrical – DC/Low Frequency
Parameter/ Equipment |
Range |
Calibration and Measurement Capability [Expressed as Uncertainty (±)] |
Reference Standard or Equipment |
AC Voltage – Generate3 |
(22 to 220) V (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz 500 kHz to 1 MHz |
0.51 mV/V + 5 mV 0.4 mV/V + 2 mV 0.38 mV/V + 0.7 mV 0.39 mV/V + 1.2 mV 0.43 mV/V + 3 mV 1.4 mV/V + 20 mV 6.3 mV/V + 50 mV 12 mV/V + 0.1 V |
Fluke 5720A |
AC Voltage – Generate3 |
(220 to 1 100) V 40 Hz to 1 kHz (1 to 20) kHz (20 to 30) kHz (220 to 750) V (30 to 50) kHz (50 to 100) kHz |
0.38 mV/V + 0.7 mV 0.39 mV/V + 1.2 mV 0.43 mV/V + 3 mV 0.39 mV/V + 1.2 mV 0.43 mV/V + 3 mV |
Fluke 5720A/5725A |
AC Voltage – Generate3 |
(1 to 5) kV (50, 60) Hz |
2.4 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 10 V |
Associated Research 5560 DT |
AC Voltage – Measure3 |
Up to 10 mV (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz 10 mV to 100 mV (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz 300 kHz to 1 MHz (1 to 2) MHz (0.1 to 1) V (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz 300 kHz to 1 MHz (1 to 2) MHz (1 to 10) V (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz 300 kHz to 1 MHz (1 to 2) MHz |
0.044 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.003 mV 0.026 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.0011 mV 0.044 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.0011 mV 0.11 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.0011 mV 0.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.0011 mV 4 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.002 mV 0.019 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.004 mV 0.019 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.002 mV 0.027 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.002 mV 0.045 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.002 mV 0.09 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.002 mV 0.31 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.01 mV 1 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.01 mV 1.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.01 mV 0.019 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.04 mV 0.019 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.02 mV 0.027 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.02 mV 0.045 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.02 mV 0.09 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.02 mV 0.31 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.1 mV 1 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.1 mV 1.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.1 mV 0.019 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.0004 V 0.019 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.0002 V 0.027 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.0002 V 0.045 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.0002 V 0.09 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.0002 V 0.31 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.001 V 1 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.001 V 1.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.001 V |
HP 3458A |
Version 001 Issued: 10/14/2016 Page 5 of 11
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org
Electrical – DC/Low Frequency
Parameter/ Equipment |
Range |
Calibration and Measurement Capability [Expressed as Uncertainty (±)] |
Reference Standard or Equipment |
AC Voltage – Measure3 |
(10 to 100) V (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (100 300 kHz to 1 MHz to 700) V (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz |
0.026 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.002 V 0.041 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.002 V 0.038 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.002 V 0.048 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.002 V 0.13 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.002 V 0.4 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.01 V 1.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.01 V 0.05 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.04 V 0.05 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.02 V 0.07 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.02 V 0.13 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.02 V 0.3 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.02 V |
HP 3458A |
AC Voltage – Measure3 |
(1 to 42) kV (50 to 60) Hz |
6.3 V |
Ross VD60 |
AC Current – Generate3 |
Up to 200 µA (10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (0.2 to 2.2) mA (10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (2.2 to 22) mA (10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (22 to 200) mA (10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (0.2 to 2.2) A 20 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz |
0.36 mA/A + 20 nA 0.25 mA/A + 12 nA 0.19 mA/A + 10 nA 0.42 mA/A + 15 nA 1.6 mA/A + 80 nA 0.36 mA/A + 50 nA 0.25 mA/A + 40 nA 0.18 mA/A + 40 nA 0.3 mA/A + 0.18 µA 1.6 mA/A + 0.8 µA 0.26 mA/A + 0.5 µA 0.24 mA/A + 0.4 µA 0.18 mA/A + 0.4 µA 0.3 mA/A + 0.7 µA 1.6 mA/A + 6 µA 0.36 mA/A + 5 µA 0.24 mA/A + 4 µA 0.18 mA/A + 3 µA 0.3 mA/A + 4 µA 1.6 mA/A + 12 µA 0.28 mA/A + 40 µA 0.6 mA/A + 0.1 mA 9.3 mA/A + 0.2 mA |
Fluke 5720A |
AC Current – Generate3 |
(2.2 to 11) A 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz |
0.81 mA/A + 0.17 mA 1.3 mA/A + 0.38 mA 4.3 mA/A + 0.75 mA |
Fluke 5720A/5725A |
AC Current – Generate3 |
(20.5 to 1 000) A (45 to 440) Hz |
1.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.9 A |
Fluke 5522A / coil |
Version 001 Issued: 10/14/2016 Page 6 of 11
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org
Electrical – DC/Low Frequency
Parameter/ Equipment |
Range |
Calibration and Measurement Capability [Expressed as Uncertainty (±)] |
Reference Standard or Equipment |
AC Current – Measure3 |
Up to 100 μA (10 to 20) Hz (20 to 45) Hz 45 Hz to 1 kHz (0.1 to 100) mA (10 to 20) Hz (20 to 45) Hz (45 to 100) Hz 100 Hz to 5 kHz (5 to 20) kHz (20 to 50) kHz (50 to 100) kHz (0.1 to 1) A (10 to 20) Hz (20 to 45) Hz (45 to 100) Hz 100 Hz to 5 kHz (5 to 20) kHz (20 to 50) kHz |
0.47 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.03 pA 0.19 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.03 pA 0.084 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.03 pA 0.47 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 20 μA 0.18 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 20 μA 0.09 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 20 μA 0.042 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 20 μA 0.078 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 20 μA 0.47 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 40 μA 0.64 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.15 mA 0.47 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.2 mA 0.19 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.2 mA 0.10 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.2 mA 0.12 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.2 mA 0.35 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.2 mA 1.2 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.4 mA |
HP 3458A |
AC Current – Measure3 |
(1 to 50) A (45 to 400) Hz |
0.06 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} |
Current shunt |
Low Frequency Power – Generate3 (45 to 65) Hz 1 PF |
Up to 20 kW |
0.42 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of rdg |
Fluke 5522A |
Electrical Calibration of Thermocouple Indicators3 – Type J Type K Type N Type T |
(-210 to 1 200) °C (-200 to 1 372) °C (-200 to 1 300) °C (-250 to 400) °C |
0.34 °C 0.48 °C 0.47 °C 0.74 °C |
Fluke 5522A |
Electrical Calibration of RTD Indicators2,3 – Pt 385, 100 Ω Pt 3926, 100 Ω Pt 3916, 100 Ω Pt 385, 200 Ω PtNi 385, 120 Ω (Ni120) Cu 427, 10 Ω [3] |
(-200 to 800) °C (-200 to 800) °C (-200 to 630) °C (-200 to 800) °C (-80 to 260) °C (100 to 260) °C |
(0.08 + 0.00021T) °C (0.08 + 0.00011T) °C (0.12 + 0.00005T) °C (0.10 + 0.00022T) °C (0.11 + 0.00023T) °C 0.35 °C |
Fluke 5522A |
Version 001 Issued: 10/14/2016 Page 7 of 11
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org
Electrical – DC/Low Frequency
Parameter/ Equipment |
Range |
Calibration and Measurement Capability [Expressed as Uncertainty (±)] |
Reference Standard or Equipment |
Oscilloscopes3 – Vertical Deflection 1 kHz Square Wave into a 50 Ω load 1 kHz Square Wave into a 1 MΩ load Rise Time Flatness Leveled Sine Wave 5 mV to 5.5 V Reference at 50 kHz Time Interval2 |
1 mV to 6.6 Vpk – pk 1 mV to 130 Vpk – pk >300 ps 50 kHz to 100 MHz (100 to 300) MHz (300 to 600) MHz (600 to 1 100) MHz 1 ns to 20 ms 50 ms to 5 s |
0.89 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 40 μV 0.35 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 40 μV (+15 / -120) ps 2.3 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.1 mV 2.6 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.1 mV 4.9 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.1 mV 6.1 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} + 0.1 mV 3.9 μs/s (29 + 1 000t) μs/s |
Fluke 5522A/SC1100 |
Electrical – RF/Microwave
Parameter/ Equipment |
Range |
Calibration and Measurement Capability [Expressed as Uncertainty (±)] |
Reference Standard or Equipment |
Attenuation3 – Measure 10 MHz to 50 GHz 10 MHz to 50 GHz 10 MHz to 50 GHz 10 MHz to 50 GHz 10 MHz to 50 GHz 10 MHz to 50 GHz 10 MHz to 50 GHz 10 MHz to 45 GHz 10 MHz to 41 GHz 10 MHz to 31.15 GHz 10 MHz to 26.5 GHz 10 MHz to 3.05 GHz |
(-10 to 0) dB (-20 to -11) dB (-30 to -21) dB (-40 to -31) dB (-50 to -41) dB (-60 to -51) dB (-70 to -61) dB (-80 to -71) dB (-90 to -81) dB (-100 to -91) dB (-110 to -101) dB (-120 to -111) dB |
0.025 dB 0.025 dB 0.025 dB 0.025 dB 0.025 dB 0.043 dB 0.043 dB 0.043 dB 0.043 dB 0.043 dB 0.056 dB 0.057 dB |
Agilent E4448A with N5532S |
RF Power – Measure 100 kHz to 4.2 GHz 10 MHz to 18 GHz 10 MHz to 26.5 GHz (26.5 to 50) GHz |
(-30 to 20) dBm (-20 to 30) dBm (-10 to 20) dBm |
4.2 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of rdg 4.9 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of rdg 4.8 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of rdg 7 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of rdg |
HP 8482A HP 8481A Agilent E4448A with N5532S HP 8487A/HP E4418B |
RF Power – Generate 10 MHz to 40 GHz 10 MHz to 40 GHz 10 MHz to 40 GHz 1.5 MHz to 1.0 GHz |
(-10 to 10) dBm (-60 to -10) dBm (-110 to -60) dBm (0 to 50) W |
1.8 dBm 2.2 dBm 2.8 dBm 1.8 mW |
HP 83460B opt 01 ENI 550 Amp, ENI 6100L Amp & HP 8482A |
Version 001 Issued: 10/14/2016 Page 8 of 11
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org
Electrical – RF/Microwave
Parameter/ Equipment |
Range |
Calibration and Measurement Capability [Expressed as Uncertainty (±)] |
Reference Standard or Equipment |
Amplitude Modulation3 – Measure 100 kHz to 10 MHz 10 MHz to 3 GHz 10 MHz to 3 GHz (3 to 26.5) GHz (3 to 26.5) GHz (26.5 to 31.5) GHz (26.5 to 31.5) GHz (31.5 to 50) GHz (31.5 to 50) GHz |
(5 to 99) {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443}Depth (5 to 20) {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443}{3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443}Depth (20 to 99) {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443}Depth (5 to 20) {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443}Depth (20 to 99) {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} Depth (5 to 20) {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} Depth (20 to 99) {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} Depth (5 to 20) {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} Depth (20 to 99) {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} Depth |
1 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} Depth 2.9 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} Depth 0.82 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} Depth 5.2 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} Depth 1.8 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} Depth 7.9 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} Depth 2.3 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} Depth 30 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} Depth 7 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} Depth |
Agilent E4448A with N5532S |
Frequency Modulation3 – Measure 250 kHz to 10 MHz 10 MHz to 6.6 GHz (6.6 to 13.2) GHz (13.2 to 31.15) GHz (31.15 to 50) GHz |
20 Hz to 10 kHz (50 to 200) Hz |
1.8 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} 1.8 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} 2.9 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} 4.4 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} 9.8 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} |
Agilent E4448A with N5532S |
Phase Modulation3 – |
100 kHz to 50 GHz |
3.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} |
Agilent E4448A with N5532S |
Audio Distortion |
400 Hz and 1 kHz |
5.8 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} |
HP 8902A |
Amplitude Modulation Distortion3 – Measure |
(0.1 to 10) MHz 10 MHz to 26.5 GHz (26.5 to 50) GHz |
1.1 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} 1.3 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} 7.2 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} |
Agilent E4448A with N5532S |
Frequency Modulation Distortion3 – Measure |
1 MHz to 50 GHz |
0.67 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} |
Agilent E4448A with N5532S |
Harmonic Distortion 20 Hz to 20 kHz 20 kHz to 50 GHz |
(-120 to 0) dB |
1.7 dB 0.91 dB |
HP 8903A Agilent E4448A with N5532S |
Power Meters3 |
3 µW to 100 mW |
0.3 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of rdg |
HP 11683A |
Mechanical
Parameter/ Equipment |
Range |
Calibration and Measurement Capability [Expressed as Uncertainty (±)] |
Reference Standard or Equipment |
Scales & Balances2,3 |
(1 to 20) g (20 to 200) g Up to 450 lb |
0.084 mg + 0.6R 0.67 mg + 0.6R 8.5 g + 0.6R |
Class 1 weights Class F weights |
Low Pressure – Measure3 |
(0 to 23) psia (0 to 7) psig |
0.02 psia 0.006 psig |
Paroscientific 760-23A |
Pressure Gauges3 |
(5 to 15 000) psi |
0.04 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of rdg |
Ametek type T |
Vacuum Gages |
(0 to 30) inHg |
0.075 inHg |
Paroscientific 760-23A |
Version 001 Issued: 10/14/2016 Page 9 of 11
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org
Mechanical
Parameter/ Equipment |
Range |
Calibration and Measurement Capability [Expressed as Uncertainty (±)] |
Reference Standard or Equipment |
Torque Wrenches3 |
(4 to 1 000) lbf·in (25 to 1 000) lbf·ft |
0.4 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} 0.49 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} |
CDI torque system |
Torque Analyzers |
Up to 250 lbf·ft |
0.11 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} |
Weights and arm |
Force Gages – Tension & Compression3 |
10 mgf to 540 lbf |
0.06 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of rdg |
Class 1 and Class F weights |
Mass – Class F |
(5 to 225) g (2 to 10) lb (25 to 50) lb |
0.49 mg 0.028 g 0.3 g |
Master balances |
Indirect Verification of Rockwell Hardness Testers3 |
(20 to 69) HRA (70 to 79) HRA (80 to 86) HRA (0 to 59) HRBW (60 to 79) HRBW (80 to 100) HRBW (20 to 39) HRC (40 to 59) HRC (60 to 70) HRC |
1.3 HRA 1.3 HRA 0.73 HRA 1.9 HRBW 1.3 HRBW 1.3 HRBW 1.3 HRC 1.3 HRC 0.73 HRC |
Hardness blocks |
Direct Verification of Durometers – Scale Accuracy Type A, B, C, D, M Indenter Geometry Length Diameter Angle |
(0 to 100) duros 0.1 in 0.05 in (30 to 35) ° |
0.01 duros 340 µin 340 µin 0.085 ° |
Master balance Optical comparator |
Thermodynamic
Parameter/ Equipment |
Range |
Calibration and Measurement Capability [Expressed as Uncertainty (±)] |
Reference Standard or Equipment |
Temperature – Measure3 |
(-200 to 600) ºC |
0.025 °C |
Hart 1502 with Fluke 5628 |
Temperature – Measuring Equipment |
(-25 to 0) °C (0 to 350) ºC |
0.16 °C 0.14 °C |
Hart 1502 with Fluke 5628 and dry block |
Humidity – Measure3 |
(0 to 90) {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} RH |
1.7 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} RH |
Vaisala HM141/HMP46 |
IR Thermometry3 |
(Ambient to 100) ºC (100 to 300) ºC (300 to 500) ºC |
0.68 ºC 0.85 ºC 1 ºC |
Fluke 9132 |
Version 001 Issued: 10/14/2016 Page 10 of 11
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.anab.org
Time and Frequency
Parameter/ Equipment |
Range |
Calibration and Measurement Capability [Expressed as Uncertainty (±)] |
Reference Standard or Equipment |
Frequency – Measuring Equipment3 |
100 MHz to 40 GHz |
2.3 parts in 109 |
HP 83640B |
Frequency – Measure |
10 MHz |
1 part in 1011 |
HP 58503A |
Frequency – Measure3 |
10 Hz to 46 GHz |
1.2 parts in 108 |
HP 53152A |
Notes:
1. Calibration and Measurement Capabilities (Expanded Uncertainties) are based on approximately a 95{3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} confidence interval, using a coverage of k=2.
2. L = length in inches, D = diameter in inches, t = time in seconds, T = temperature in degree C, R = resolution of the device under test.
3. This laboratory offers field calibration services.
4. {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} = percent of reading unless otherwise indicated.
5. This scope is formatted as part of a single document including Certificate of Accreditation No. AC-2080.01
Version 001 Issued: 10/14/2016 Page 11 of 11