SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017, ANSI/NCSL Z540-1-1994 (R2002) AND ANSI/NCSL Z540.3-2006 (R2013)
Technical Maintenance, Inc.
425 Hayden Station Road,
Suite B
Windsor, CT 06095
Scott Chamberlain 860-219-0046
CALIBRATION
Valid to: September 20, 2019 Certificate Number: AC-2080.04
Electrical – DC/Low Frequency
Range |
Expanded Uncertainty of Measurement (+/-) |
Reference Standard, Method, and/or Equipment |
|
DC Voltage1 – Generate |
Up to 330 mV 330 mV to 3.3 V (3.3 to 33) V (33 to 330) V (100 to 1 000) V |
24 µV/V + 1 µV 14 µV/V + 2 µV 15 µV/V + 20 µV 22 µV/V + 0.15 mV 22 µV/V + 1.5 mV |
Fluke 5522A |
DC Voltage1– Measure |
Up to 100 mV 100 mV to 1 V (1 to 10) V (10 to 100) V (100 to 1 000) V |
17 µV/V + 0.3 µV 12 µV/V + 0.3 µV 11 µV/V + 0.5 µV 52 μV/V + 30 µV 61 μV/V + 0.1 mV |
HP 3458A |
(1 to 60) kV |
0.12 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading |
Ross VD60-6.2Y-A-LB- AL |
|
DC Current – Generate1 |
Up to 330 µA 330 µA to 3.3 mA (3.3 to 33) mA (33 to 330) mA 330 mA to 1.1 A (1.1 to 3) A (3 to 11) A (11 to 20.5) A |
0.18 mA/A + 0.02 µA 0.12 mA/A + 0.05 µA 0.12 mA/A + 0.25 µA 0.24 mA/A + 2.5 µA 0.24 mA/A + 40 μA 0.44 mA/A + 40 μA 0.6 mA/A + 0.5 mA 1.6 mA/A + 0.75 mA |
Fluke 5522A |
Parameter/Equipment |
Range |
Expanded Uncertainty of Measurement (+/-) |
Reference Standard, Method, and/or Equipment |
DC Current – Measure |
Up to 100 nA 100 nA to 1 µA (1 to 10) µA (10 to 100) µA 100 µA to 3mA (10 to 100) mA 1 mA to 1 A |
35 µA/A + 0.04 nA 24 µA/A + 0.04 nA 24 µA/A + 0.1 nA 24 µA/A + 0.8 nA 24 µA/A + 0.05 nA 42 µA/A + 0.5 µA 0.15 mA/A + 10 µA |
HP 3458A |
AC Voltage – Generate1 |
(1 to 33) mV (10 to 45) Hz 45 Hz to 10 kHz (10 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 500) kHz (33 to 330) mV (10 to 45) Hz 45 Hz to 10 kHz (10 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 500) kHz 330 mV to 3.3 V (10 to 45) Hz 45 Hz to 10 kHz (10 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 500) kHz |
0.95 mV/V + 6 μV 0.19 mV/V + 6 μV 0.24 mV/V + 6 μV 1.2 mV/V + 6 μV 4 mV/V + 12 μV 9.3 mV/V + 50 μV 0.35 mV/V + 8 μV 0.17 mV/V + 8 μV 0.19 mV/V + 8 μV 0.41 mV/V + 8 μV 0.93 mV/V + 32 μV 2.3 mV/V + 70 μV 0.35 mV/V + 50 μV 0.18 mV/V + 60 μV 0.22 µV/V + 60 μV 0.35 mV/V + 50 μV 0.81 mV/V + 0.13 mV 2.8 mV/V + 0.6 mV |
Fluke 5522A |
Parameter/Equipment |
Range |
Expanded Uncertainty of Measurement (+/-) |
Reference Standard, Method, and/or Equipment |
AC Voltage – Generate1 |
(3.3 to 33) V (10 to 45) Hz 45 Hz to 10 kHz (10 to 20) kHz (20 to 50) kHz (50 to 100) kHz (33 to 330) V 45 Hz to 1 kHz (1 to 10) kHz (10 to 20) kHz (20 to 50) kHz (50 to 100) kHz (330 to 1 020) V 45 Hz to 10 kHz |
0.36 mV/V + 0.65 mV 0.2 mV/V + 0.6 mV 0.47 mV/V + 0.6 mV 0.42 mV/V + 0.6 mV 1 mV/V + 1.6 mV 0.85 mV/V + 2 mV 0.85 mV/V + 6 mV 0.87 mV/V + 6 mV 0.9 mV/V + 6 mV 2.5 mV/V + 50 mV 8.2 mV/V + 10 mV |
Fluke 5522A |
AC Voltage – Measure |
Up to 10 mV (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz 10 mV to 100 mV (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz 300 kHz to 1 MHz (1 to 2) MHz 100 mV to 1 V (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz 300 kHz to 1 MHz (1 to 2) MHz |
0.044 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.003 mV 0.026 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.0011 mV 0.044 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.0011 mV 0.11 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.0011 mV 0.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.0011 mV 4 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 mV 0.019 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.004 mV 0.019 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 mV 0.027 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 mV 0.045 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 mV 0.09 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 mV 0.31 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.01 mV 1 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.01 mV 1.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.01 mV 0.019 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.04 mV 0.019 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.02 mV 0.027 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.02 mV 0.045 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.02 mV 0.09 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.02 mV 0.31 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.1 mV 1 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.1 mV 1.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.1 mV |
HP 3458A |
Parameter/Equipment |
Range |
Expanded Uncertainty of Measurement (+/-) |
Reference Standard, Method, and/or Equipment |
AC Voltage – Measure |
1 V to 10 V (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz 300 kHz to 1 MHz (1 to 2) MHz (10 to 100) V (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz 300 kHz to 1 MHz (100 to 700) V (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz |
0.19 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.4 mV 0.019 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mV 0.027 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mV 0.045 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mV 0.09 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mV 0.31 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 1 mV 1 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 1 mV 1.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 1 mV 0.026 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 V 0.041 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 V 0.038 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 V 0.048 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 V 0.13 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.002 V 0.4 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.01 V 1.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.01 V 0.05 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.04 V 0.05 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.02 V 0.07 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.02 V 0.13 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.02 V 0.3 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.02 V |
HP 3458A |
(1 to 42) kV Up to 60 Hz |
0.6 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading |
Ross VD60-6.2Y-A-LB- AL |
|
AC Current – Generate1 40 Hz to 1 kHz |
(29 to 330) µA 330 μA to 3.3 mA (3.3 to 33) mA (33 mA to 330) mA 330 mA to 1.1 A (1.1 to 3) A (3 to 11) A (11 to 20.5) A |
0.22 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.1 μA 0.13 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.15 μA 0.08 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 2 μA 0.08 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 20 μA
0.13 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 2 mA 0.18 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 5 mA |
Fluke 5522A |
Parameter/Equipment |
Range |
Expanded Uncertainty of Measurement (+/-) |
Reference Standard, Method, and/or Equipment |
AC Current – Measure |
Up to 100 μA (10 to 20) Hz (20 to 45) Hz 45 Hz to 1 kHz (0.1 to 1) mA (10 to 20) Hz (20 to 45) Hz (45 to 100) Hz 100 Hz to 5 kHz (5 to 20) kHz (20 to 50) kHz (50 to 100) kHz (1 to 10) mA (10 to 20) Hz (20 to 45) Hz (45 to 100) Hz 100 Hz to 5 Hz (5 to 20) kHz (20 to 50) kHz (10 to 100) mA (10 to 20) Hz (20 to 45) Hz (45 to 100) Hz 100 Hz to 5 Hz (5 to 20) kHz (20 to 50) kHz (0.1 to 1) A (10 to 20) Hz (20 to 45) Hz (45 to 100) Hz 100 Hz to 5 Hz (5 to 20) kHz (20 to 50) kHz |
0.41 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.03 pA 0.16 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.03 pA 0.07 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.03 pA 0.41 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mA 0.16 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mA 0.069 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mA 0.038 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mA 0.069 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mA 0.41 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mA 0.56 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 1.5 mA 0.41 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 2 mA 0.16 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 2 mA 0.069 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 2 mA 0.038 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 2 mA 0.41 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 4 mA 0.56 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 15 mA 0.41 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 20 mA 0.16 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 20 mA 0.069 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 20 mA 0.038 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 20 mA 0.41 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 40 mA 0.56 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 150 mA 0.41 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mA 0.16 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mA 0.069 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mA 0.038 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mA 0.31 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.2 mA 1 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.4 mA |
HP 3458A |
Resistance – Generate1 |
Up to 11 Ω (11 to 33) Ω (33 to 110) Ω (110 to 330) Ω 330 Ω to 1.1 kΩ |
51 μΩ/Ω + 0.01 Ω 42μΩ/Ω + 0.015 Ω 34 μΩ/Ω + 0.015 Ω 33 μΩ/Ω + 0.02 Ω 34 μΩ/Ω + 0.02 Ω |
Fluke 5522A |
Parameter/Equipment |
Range |
Expanded Uncertainty of Measurement (+/-) |
Reference Standard, Method, and/or Equipment |
Resistance – Generate1 |
(1.1 to 3.3) kΩ (3.3 to 11) kΩ (11 to 33) kΩ (33 to 110) kΩ (110 to 330) kΩ 0.33 MΩ to 1.1 MΩ (1.1 to 3.3) MΩ (3.3 to 11) MΩ (11 to 33) MΩ (33 to 110) MΩ (110 to 330) MΩ 330 MΩ to 1.1 GΩ |
34 μΩ/Ω + 0.2 Ω 34 μΩ/Ω + 0.1 Ω 34 μΩ/Ω + 1 Ω 34 μΩ/Ω + 1 Ω 41 μΩ/Ω + 10 Ω 39 μΩ/Ω + 10 Ω 71 μΩ/Ω + 150 Ω 150 μΩ/Ω + 250 Ω 300 μΩ/Ω + 2.5 kΩ 610 μΩ/Ω + 3 kΩ 3.5 mΩ/Ω + 0.1 MΩ 17 mΩ/Ω + 0.5 MΩ |
Fluke 5522A |
Resistance – Measure |
Up to 10 Ω (10 to 100) Ω 100 Ω to 1 kΩ (1 to 10) kΩ (10 to 100) kΩ 100 kΩ to 1MΩ (1 to 10) MΩ (10 to 100) MΩ 100 MΩ to 1 GΩ |
27 μΩ/Ω + 50 μΩ 18 μΩ/Ω + 0.5 mΩ 15 μΩ/Ω + 0.5 mΩ
26 μΩ/Ω + 2 Ω 70 μΩ/Ω + 100 Ω 590 μΩ/Ω + 1 kΩ 0.78 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 10 kΩ |
HP 3458A |
Electrical Calibration of Thermocouple Indicating Devices1 – |
Type E (230 to 1 000) ºC Type J (-210 to 1 200) ºC Type K (-200 to 1 372) ºC Type T (-250 to 400) ºC Type R (0 to 1 767) ºC Type S (0 to 1 767) ºC |
0.60 ºC 0.5 ºC 0.49 ºC 0.74 ºC 0.76 ºC 0.63 ºC |
Fluke 5522A |
Parameter/Equipment |
Range |
Expanded Uncertainty of Measurement (+/-) |
Reference Standard, Method, and/or Equipment |
Capacitance – Generate |
(0.19 to 0.1099) nF (1.1 to 3.2999) nF (3.3 to 329.999) nF (0.33 to 1.09999) µF (1.1 to 3.29999) µF (3.3 to 10.9999) µF (11 to 32.999) µF (33 to 109.999) µF (110 to 329.999) µF (0.33 to 1.09999) mF (1.1 to 3.2999) mF (3.3 to 10.9999) mF (11 to 32.9999) mF (33 to 110) mF |
1.2 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.01 nF 1.2 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.01 nF 0.38 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.3 nF 0.37 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 3 nF
0.53 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 30 nF 0.58 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.1 µF 0.58 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.3 µF
0.65 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 30 μF 1.3 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.1 mF |
Fluke 5522A |
Oscilloscopes1,2– Amplitude – Square Wave (peak to peak) 50 Ω 1 MΩ Rise Time Flatness (50 kHz reference) Time Marker |
1 mV to 6.6 V 1 mV to 130 V <350 ps 5 mV to 5.5 V 50 kHz to 100 MHz (100 to 300) MHz (300 to 600) MHz (600 to 1 100) MHz 1 nS to 20 mS 50 mS to 5 S |
0.25 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 40 µV 0.1 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 40 µV + 13 ps / -0.120 ps 2.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.1 mV 3.4 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.1 mV 5.5 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.1 mV 6.9 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading + 0.1 mV 2.7 parts in 106 S (25 + 1 000t) parts in 106 S |
Fluke 5522A /SC1100 |
Length – Dimensional metrology
Range |
Expanded Uncertainty of Measurement (+/-) |
Reference Standard, Method, and/or Equipment |
|
Gage Blocks2 |
Up to 8 in |
(5.2+ 2.3L) µin |
Master gage blocks, P&W UMM |
Parameter/Equipment |
Range |
Expanded Uncertainty of Measurement (+/-) |
Reference Standard, Method, and/or Equipment |
Micrometers1,2 |
Up to 18 in |
(29+ 5L) µin |
Gage blocks |
Laser Micrometer1 |
(0.5 to 1.5) in |
170 µin |
Pin Gages |
Calipers1,2 |
Up to 20 in |
(30 + 14L) µin |
Gage blocks |
Dial Indicators1 |
Up to 1 in |
350 µin |
Gage blocks |
Cylindrical Gages2 – Rings Plain Pins, Plugs |
(0.02 to 8) in (0 to 8) in |
(6.5 + 1.6D) µin (3 + 7D) µin |
Master gage blocks, P&W UMM |
Linear Scales – Rulers1 |
Up to 36 in |
0.013 in |
Gage blocks |
Pi Tapes1 |
Up to 3 in |
450 μin |
Plug Gages |
Tape Measures1 |
Up to 150 ft |
0.19 in |
Master tape |
Microscopes1 |
Up to 2 in |
100 µin |
Stage Micrometer |
Mass
Range |
Expanded Uncertainty of Measurement (+/-) |
Reference Standard, Method, and/or Equipment |
|
Scales & Balances1 |
(1 to 20) g (20 to 200) g (200 to 500) g |
0.16 mg + 0.6R 1.4 mg + 0.6R 31 mg + 0.6R |
ASTM E617 Class 1 weights |
Up to 400 lb |
21 g + 0.6R |
NIST Class F weights |
|
Pressure1 |
(-14 to 0) psig Up to 1 000 psig |
1.4 psig 1 psig |
Fluke 700RG08 pressure gauge calibrator |
Torque Wrenches1 |
4 lbfin to 250 lbfft (250 to 1 000) lbfft |
0.9 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading 1.3 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443} of reading |
CDI 5000 ST torque tester |
Range |
Expanded Uncertainty of Measurement (+/-) |
Reference Standard, Method, and/or Equipment |
|
Relative Humidity Generate |
(10 to 98) {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443}RH |
0.51 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443}RH |
Thunder Scientific 2500 |
Relative Humidity Measure |
(0 to 90) {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443}RH |
1.7 {3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443}RH |
Vaisala M170/HMP75 |
Temperature – Measure1 |
(-25 to 0) °C (0 to 140) °C (140 to 600) °C |
2.8 °C 0.21 °C 1.6 °C |
Fluke 714 & 80PK-9 Fluke 2180A Fluke 714 & 80PK-9 |
Time and Frequency
Range |
Expanded Uncertainty of Measurement (+/-) |
Reference Standard, Method, and/or Equipment |
|
Frequency – Measure |
10 MHz |
1 x 10-11 Hz/Hz |
Agilent 58503A/B GPS |
Frequency – Measure1 |
0 to 100 MHz |
2.1 x 10-7 Hz/Hz |
HP 5334A |
Calibration and Measurement Capability (CMC) is expressed in terms of the measurement parameter, measurement range, expanded uncertainty of measurement and reference standard, method, and/or equipment. The expanded uncertainty of measurement is expressed as the standard uncertainty of the measurement multiplied by a coverage factor of 2 (k=2), corresponding to a confidence level of approximately 95{3abc3950d4f1f8f388f3ff046375d7d82c013b00bce2a30b1200ecdfc3f73443}.
Notes:
-
On-site calibration service is available for this parameter, since on-site conditions are typically more variable than those in the laboratory, larger measurement uncertainties are expected on-site than what is reported on the accredited scope.
-
L = length in inches, D = diameter in inches, R = resolution of device under test, t = time in seconds.
-
This scope is formatted as part of a single document including Certificate of Accreditation No. AC-2080.04.